10 Transport, packaging, storage
The instrument can be damaged or destroyed if shipped incorrectly. To transport or store the
instrument, please use the original box or a box padded with sufficient shock-absorbing material.
For storage in humid areas or shipment overseas, the device should be placed in welded foil
(ideally along with silica gel) to protect it from humidity.
The pyrometer is designed for a storage temperature of -20 to 70°C with non-condensing condi-
tions. Other kind of storage can damage or malfunction the pyrometer.
11 Maintenance
The pyrometer does not have any parts which require regular service, only the lens has to be kept
clean. The lens can be cleaned with a soft cloth in combination with alcohol (do not use acid
solutions or dilution). It is also possible to use standard cleaning tissue for glasses or camera
lenses.
It is recommended to send the pyrometer every two years to LumaSense for recalibration.
12 Trouble shooting
Before sending the pyrometer for repair, try to find the error and to solve the problem with the help
of the following list.
Temperature indication too low
Incorrect alignment of the pyrometer to the object
New correct alignment to achieve the max. temperature signal
Measuring object is smaller than spot size
Choose correct measuring distance
Measuring object is not always in the measuring spot of the pyrometer (e.g. swinging wire or
pouring stream)
Use max. value storage
Emissivity set too high
Set lower correct emissivity corresponding to the material
Lens contaminated or scratched
Clean lens carefully
Temperature indication too high
Emissivity set too low
Set higher correct emissivity corresponding to the material
The measurement is influenced by reflections of hotter machine parts
Try to avoid the influence of the interfering radiation or change measuring position
Measuring errors
Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens
Clean lens
Air contamination in the sighting path between pyrometer and object
Change position of the pyrometer with a clean sighting path
Strong HF-interferences
Change position of the pyrometer