M390S Manual Operation 18
Typical Radiance temperature calibration look-up table (LUT) used
To gain access to the custom look-up table:
Refer to the 4 elliptical buttons on the face of the controller and use this procedure.
1. Press and Hold 1 until ACCESS appears. If Access Level and Level 1 appears, skip to
step 3.
2. Press 4 until Level 3 appears.
3. Press 4 until 3 appears. Wait and "pass" appears.
4. Press 1 until Lin16 appears.
5. Press 2. until display shows the Output # that corresponds to the calibration point
shown in the listing on the previous page.
6. Use UP/DWN buttons (3 and 4) to dial in the measured radiance temperature from a
freezing point transfer device or other radiance standard. Adjust Out# only!
7. Press 2 to view/change more correction points.
8. When finished with all calibrations and to lock the controller, press 1 until ACCESS
appears.
9. Press 3 until Level 1 appears.
1 2 3